Reference dose levels for dental panoramic radiography in Gwangju, South Korea.

Authors
  • Lee, Jae-Seo1
  • Kim, Young-Hee
  • Yoon, Suk-Ja
  • Kang, Byung-Cheol
  • 1 Department of Oral and Maxillofacial Radiology, School of Dentistry, Dental Science Research Institute, Chonnam National University, 77 Yongbong-ro Buk-gu, Gwangju 500-757, South Korea. , (North Korea)
Type
Published Article
Journal
Radiation Protection Dosimetry
Publisher
Oxford University Press
Publication Date
Dec 01, 2010
Volume
142
Issue
2-4
Pages
184–190
Identifiers
DOI: 10.1093/rpd/ncq189
PMID: 20702464
Source
Medline
License
Unknown

Abstract

This study assessed the reference dose levels for dental panoramic radiography in Gwangju city, South Korea based on the dose width product (DWP) and compared them with those already established elsewhere. A total of 44 panoramic dental radiographic sets (36 digital and 8 analogue panoramic sets) in 41 dental clinics in Gwangju city were chosen. The third quartile DWP was determined from 429 surface dose measurements of the adult surface dose in panoramic dental radiography. The third quartile DWP for panoramic radiography was 60.1 mGy mm. The proposed DWP reference levels of 60.1 mGy mm were less than or equal to those previously reported in other countries, such as Italy and UK, and acceptable for panoramic radiography in Gwangju, South Korea.