Quality Control Strategy for Automated CBC: A Laboratory Point of View Deducted from an Internal Study Organised in an Emergency Laboratory

Authors
  • Oprea, Oana Roxana1,
  • Hutanu, Adina1, 1
  • Pavelea, Oana
  • Kodori, David Robert
  • Dobreanu, Minodora1, 1
  • 1 George Emil Palade University of Medicine, Pharmacy, Science, and Technology of Targu Mures, Romania , (Romania)
Type
Published Article
Journal
Revista Romana de Medicina de Laborator
Publisher
De Gruyter Open Sp. z o.o.
Publication Date
Jan 01, 2020
Volume
28
Issue
1
Pages
19–27
Identifiers
DOI: 10.2478/rrlm-2020-0012
Source
De Gruyter
Keywords
License
Green

Abstract

Introduction: The aim of this study was to determine the performance of the total testing process of complete blood count (CBC) on two different instruments in an emergency setting of a county hospital, and to design an appropriate internal quality control plan. Materials and method: Two models of Statistical Quality Control (SQC) were evaluated on Sysmex XT-1800i and Cell-Dyne Ruby: 3 levels of commercial blood every 8 hours (N=9) and an alternative model using 3 levels every 12 hours (N=6) as shift changes. Total Error (TE) was calculated using the formula: TE=Bias%+1.65xCV%; Sigma score was calculated using the formula: Sigma=[(TEa%–Bias%]/CV%. Values for coefficient of variation (CV%) and standard deviation (SD) were obtained from laboratory data and Bias% from proficiency testing. For the pre-analytical phase Sigma score was calculated, while for post-analytical phase the turnaround time (TAT) was assessed. Results: TE for all directly measured parameters, for both instruments, had lower values than Total Error allowable (TEa). CV% for almost all parameters had lower values than CV% derived from biological variation except for platelets (PLT) at low level on Sysmex XT-1800i and red blood cells (RBC) on Cell-Dyne Ruby. Sigma score ranged from as low as 2 to 10. Sigma score for pre-analytical phase was 4.2 and turnaround time was 36 minutes on average. Conclusions: Given the performances of the total testing process implemented for CBC in our laboratory, performing the internal control after every 50 samples/batch seems to fulfill both the Health Ministry Order (HMO) 1301/2007 and International Organization for Standardization ISO 15189:2013 recommendation. All quality instruments must work together to assure better patient results and every laboratory should design its own control plan that is appropriate for better quality achievement.

Report this publication

Statistics

Seen <100 times