Performance evaluation of new Abbott Alinity hq hematology analyzer.

Authors
  • Woo, Kwang-Sook1
  • Jeong, In-Hwa1
  • An, Gyu-Dae1
  • Lim, Hyeon-Ho1
  • Han, Jin-Yeong1
  • 1 Department of Laboratory Medicine, Dong-A University College of Medicine, Busan, Korea. , (North Korea)
Type
Published Article
Journal
International journal of laboratory hematology
Publication Date
Oct 01, 2019
Volume
41
Issue
5
Pages
657–663
Identifiers
DOI: 10.1111/ijlh.13082
PMID: 31271528
Source
Medline
Keywords
Language
English
License
Unknown

Abstract

Abbott Alinity hq is a next-generation automated hematology analyzer providing complete blood count (CBC) with 6-part white blood cells (WBC) differential counts. The purpose of this study was to evaluate the performance of the analyzer to verify the diagnostic and clinical utility of the Abbott Alinity hq automated system. We evaluated specimen stability, precision, linearity, carry-over, and method comparison to assess the performance of Alinity hq. For comparison of the Alinity hq with Sysmex XN-9000, totally 314 samples from adult and pediatric patients including both normal and abnormal hematology profiles were analyzed in parallel. The Alinity hq was also compared with the manual differential counts for the same 314 samples. At 4°C, the Alinity hq analyzer showed no significant changes in CBC and WBC differential count up to 48 hours. When stored at room temperature (18-25°C), all parameters except the mean platelet volume (MPV) were stable up to 36 hours. The Abbott Alinity hq analyzer demonstrated excellent reproducibility and between-batch precision for all CBC and WBC differential parameters. WBC, red blood cells (RBC), hemoglobin (HGB), and platelets showed good linearity and acceptable carry-over. Comparison with a Sysmex XN-9000 analyzer and manual 400-cell differential showed excellent correlation for CBC and WBC differential count parameters (correlation coefficient = 0.815-0.999) except for mean corpuscular hemoglobin concentration (MCHC) and basophils. We performed initial validation studies and confirmed performance specifications on specimen stability, precision, linearity, carry-over, and method comparison. The Abbott Alinity hq analyzer showed good analytical performance for all standard CBC parameters. © 2019 John Wiley & Sons Ltd.

Report this publication

Statistics

Seen <100 times