Multiple inappropriate rate drop responses triggered by ventricular premature beats.

Authors
Type
Published Article
Journal
EP Europace
1099-5129
Publisher
Oxford University Press
Publication Date
Volume
13
Issue
7
Pages
1046–1046
Identifiers
DOI: 10.1093/europace/eur012
PMID: 21273193
Source
Medline

Abstract

We report the case of a 70-year-old patient presenting with palpitations who was found to have recurrent inappropriate rate drop responses on interrogation of his pacemaker. These were found to be triggered by interpolated premature ventricular complexes (PVCs). We discuss the drop detect algorithm and how interpolated PVCs triggered the rate drop response.