High-resolution high-speed panoramic cardiac imaging system.

Authors
  • Evertson, Dale W
  • Holcomb, Mark R
  • Eames, Matthew C
  • Bray, Mark-Anthony
  • Sidorov, Veniamin Y
  • Xu, Junkai
  • Wingard, Holley
  • Dobrovolny, Hana M
  • Woods, Marcella C
  • Gauthier, Daniel J
  • Wikswo, John P
Type
Published Article
Journal
IEEE transactions on bio-medical engineering
Publication Date
Mar 01, 2008
Volume
55
Issue
3
Pages
1241–1243
Identifiers
DOI: 10.1109/TBME.2007.912417
PMID: 18334422
Source
Medline
License
Unknown

Abstract

A panoramic cardiac imaging system consisting of three high-speed CCD cameras has been developed to image the surface electrophysiology of a rabbit heart via fluorescence imaging using a voltage-sensitive fluorescent dye. A robust, unique mechanical system was designed to accommodate the three cameras and to adapt to the requirements of future experiments. A unified computer interface was created for this application - a single workstation controls all three CCD cameras, illumination, stimulation, and a stepping motor that rotates the heart. The geometric reconstruction algorithms were adapted from a previous cardiac imaging system. We demonstrate the system by imaging a polymorphic cardiac tachycardia.