Affordable Access

High resolution gamma ray tomography scanner for flow measurement and non-destructive testing applications.

Authors
Type
Published Article
Journal
The Review of scientific instruments
Publication Date
Volume
78
Issue
10
Pages
103704–103704
Identifiers
PMID: 17979425
Source
Medline

Abstract

We report on the development of a high resolution gamma ray tomography scanner that is operated with a Cs-137 isotopic source at 662 keV gamma photon energy and achieves a spatial image resolution of 0.2 line pairs/ mm at 10% modulation transfer function for noncollimated detectors. It is primarily intended for the scientific study of flow regimes and phase fraction distributions in fuel element assemblies, chemical reactors, pipelines, and hydrodynamic machines. Furthermore, it is applicable to nondestructive testing of larger radiologically dense objects. The radiation detector is based on advanced avalanche photodiode technology in conjunction with lutetium yttrium orthosilicate scintillation crystals. The detector arc comprises 320 single detector elements which are operated in pulse counting mode. For measurements at fixed vessels or plant components, we built a computed tomography scanner gantry that comprises rotational and translational stages, power supply via slip rings, and data communication to the measurement personal computer via wireless local area network.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times