Affordable Access

Publisher Website

An assessment of Stat Tek: A reagent-strip method for glucose determination

Authors
Journal
Clinical Biochemistry
0009-9120
Publisher
Elsevier
Publication Date
Volume
13
Issue
1
Identifiers
DOI: 10.1016/s0009-9120(80)90642-6

Abstract

Stat Tek, a reagent — strip method for glucose determination, has been evaluated. Aqueous solution of glucose standard is not recommended. The linearity of serum glucose is between 630 to 3,450 mg/l. The within run precision is between 0.7% to 3.0%. The day-to-day precision is between 2.8% to 5.6%. The recovery of glucose added to serum is between 95 to 101%. The correlation of glucose by Stat-Tek and SMA 6/60 show small proportional and constant errors. Stat-Tek is simple and easy to use; however, proper quality control and training of the user are essential for the correct measurement of glucose.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times
0 Comments

More articles like this

An assessment of Stat Tek: a reagent-strip method...

on Clinical Biochemistry February 1980

Reagent-strip methods for glucose.

on The Medical journal of Austral... Jan 04, 1988

Reagent-strip method of blood glucose determinatio...

on Journal of occupational medici... October 1965

The effect of the haematocrit value on the determi...

on The Medical journal of Austral... Sep 21, 1987
More articles like this..