An EM simulation based design flow for custom-built MR coils incorporating signal and noise.

Authors
  • Horneff, Andreas1
  • Eder, Michael2
  • Hell, Erich3
  • Ulrici, Johannes3
  • Felder, Jorg4
  • Rasche, Volker2
  • Anders, Jens5
  • 1 University Hospital Ulm, Germany and Sirona Dental Systems, Bensheim, Germany. , (Germany)
  • 2 University Hospital Ulm, Germany. , (Germany)
  • 3 Sirona Dental Systems, Bensheim, Germany. , (Germany)
  • 4 Forschungszentrum Jülich, Germany. , (Germany)
  • 5 University of Ulm, Germany. , (Germany)
Type
Published Article
Journal
IEEE Transactions on Medical Imaging
Publisher
Institute of Electrical and Electronics Engineers
Publication Date
Oct 17, 2017
Pages
1–1
Identifiers
DOI: 10.1109/TMI.2017.2764160
PMID: 29053444
Source
Medline
Keywords
License
Unknown

Abstract

Developing custom-built MR coils is a cumbersome task, in which an a priori prediction of the coils' SNR performance, their sensitivity pattern and their depth of penetration helps to greatly speed up the design process by reducing the required hardware manufacturing iterations. The simulationbased design flow presented in this paper takes the entire MR imaging process into account. That is, it includes all geometric and material properties of the coil and the phantom, the thermal noise as well as the target MR sequences. The proposed simulation-driven design flow is validated using a manufactured prototype coil, whose performance was optimized regarding its SNR performance, based on the presented design flow, by comparing the coil's measured performance against the simulated results. In these experiments, the mean and the standard deviation of the relative error between the simulated and measured coil sensitivity pattern were found to be μ = 1.79% and σ = 3.15%. Moreover, the peak deviation between the simulated and measured voxel SNR was found to be less than 4%, indicating that simulations are in good accordance with the measured results, validating the proposed software-based design approach.Developing custom-built MR coils is a cumbersome task, in which an a priori prediction of the coils' SNR performance, their sensitivity pattern and their depth of penetration helps to greatly speed up the design process by reducing the required hardware manufacturing iterations. The simulationbased design flow presented in this paper takes the entire MR imaging process into account. That is, it includes all geometric and material properties of the coil and the phantom, the thermal noise as well as the target MR sequences. The proposed simulation-driven design flow is validated using a manufactured prototype coil, whose performance was optimized regarding its SNR performance, based on the presented design flow, by comparing the coil's measured performance against the simulated results. In these experiments, the mean and the standard deviation of the relative error between the simulated and measured coil sensitivity pattern were found to be μ = 1.79% and σ = 3.15%. Moreover, the peak deviation between the simulated and measured voxel SNR was found to be less than 4%, indicating that simulations are in good accordance with the measured results, validating the proposed software-based design approach.