Affordable Access

Publisher Website

A new aliasing extension method for ultrasonic 2-dimensional pulsed Doppler systems

Authors
Journal
Ultrasonic Imaging
0161-7346
Publisher
SAGE Publications
Publication Date
Volume
11
Issue
4
Identifiers
DOI: 10.1016/0161-7346(89)90077-1
Keywords
  • Mean-Frequency Aliasing
  • Pulsed Doppler System
  • Spatial Axis
  • Temporal Axis
  • Velocity Profile

Abstract

Abstract The mean frequency aliasing problem originating from the pulse repetition frequency is one of major limitations in ultrasound pulsed Doppler systems. A conventional approach to resolving this problem is to track the mean frequencies close to and beyond the Nyquist frequency along the temporal axis. In this paper, a new concept of tracking the mean frequencies along the spatial axis is proposed for the same problem. The technique is fault tolerable and more suitable for multigate and 2-D Doppler systems than conventional methods. Simulation and experimental results show that the proposed system has improved performance compared with that of the conventional systems.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times
0 Comments

More articles like this

Clinical evaluation of a new anti-aliasing techniq...

on Ultrasound in Medicine & Biolo... Jan 01, 1989

Clinical evaluation of a new anti-aliasing techniq...

on Ultrasound in Medicine & Biolo... 1989

On the aliasing problem in pulsed Doppler cardiac...

on Journal of clinical ultrasound... 1984
More articles like this..