Development of hemoglobin A1c certified reference material by liquid chromatography isotope dilution mass spectrometry.

Authors
  • Bi, Jiaming
  • Wu, Liqing
  • Yang, Bin
  • Yang, Yi
  • Wang, Jing
Type
Published Article
Journal
Analytical and Bioanalytical Chemistry
Publisher
Springer-Verlag
Publication Date
Apr 01, 2012
Volume
403
Issue
2
Pages
549–554
Identifiers
DOI: 10.1007/s00216-012-5834-9
PMID: 22349405
Source
Medline
License
Unknown

Abstract

We report the development of a National Institute of Metrology (NIM) hemoglobin A(1c) (HbA(1c)) certified reference material (CRM). Each CRM unit contains about 10 μL of hemoglobin. Both hemoglobin and glycated hemoglobin were quantitatively determined by high-performance liquid chromatography (HPLC)-isotope dilution mass spectrometry (IDMS) with synthesized VHLTPE and glycated VHLTPE as standards. The mass fraction of synthesized VHLTPE or glycated VHLTPE was also quantitatively determined by HPLC-IDMS with NIM amino acid CRMs as standards. The homogeneity and stability of the CRMs were examined with a commercial HbA(1c) analyzer based on the HPLC principle. Fifteen units were randomly selected for homogeneity examination, and statistical analysis showed there was no inhomogeneity. Examination of the stability showed that the CRM was stable for at least 6 months at -80 °C. Uncertainty components of the balance, amino acid purity, hydrolysis and proteolysis efficiency, method reproducibility, homogeneity, and stability were taken into consideration for uncertainty evaluation. The certified value of NIM HbA(1c) CRM was expressed as the ratio of HbA(1c) to total hemoglobin in moles, and was (9.6 ± 1.9)%. The CRM can be used as a calibration or validation standard for clinical diagnostics. It is expected to improve the comparability for HbA(1c) measurement in China.

Report this publication

Statistics

Seen <100 times