Affordable Access

Access to the full text

Determination of Chemically Bound Water in Metakaolin

Authors
  • Vdovin, K. N.1
  • Pivovarova, K. G.1
  • Ponamareva, T. B.1
  • Feoktistov, N. A.1
  • 1 Nosov Magnitogorsk State Technical University (NMSTU), Magnitogorsk, Russia , Magnitogorsk (Russia)
Type
Published Article
Journal
Refractories and Industrial Ceramics
Publisher
Springer US
Publication Date
Mar 01, 2019
Volume
59
Issue
6
Pages
654–657
Identifiers
DOI: 10.1007/s11148-019-00290-5
Source
Springer Nature
Keywords
License
Yellow

Abstract

The technique is presented for the determination of chemically bound water in metakaolin (with the mass fraction of water being within 0.5 and 5 %) by IR spectrometry at 900°C using the ELTRA (Germany) CW-800M infrared spectrometer. A single sample can be tested within 5 – 6 min, allowing for quality control of incoming metakaolin samples keeping defective samples out of production. The possibility of water determination at 600°C was shown, in which case analysis time increases to 20 – 22 min. The developed technique for the determination of chemically bound water in metakaolin provides results with sufficient accuracy and precision.