Affordable Access

Publisher Website

In vivolocalized1H NMR spectroscopy at 11.7 Tesla

Authors
Journal
Journal of Magnetic Resonance (1969)
0022-2364
Publisher
Elsevier
Publication Date
Volume
94
Issue
1
Identifiers
DOI: 10.1016/0022-2364(91)90299-9
Disciplines
  • Design

Abstract

Abstract The SPACE volume-selection technique has been used to acquire high-resolution 1H spectra from the brain of neonate mice at 11.7 T (500 MHz). Spectra were acquired from voxels smaller than 20 μl. The spectra display elevated intensities of resonances arising from taurine and reduced intensities of those arising from N-acetylaspartate, when compared to those of mature animals, correlating well with in vitro studies. An integrated probe design consisting of separate transmission and reception RF coils and linear gradient coils is described. Comments are made concerning the advantages and disadvantages of performing gradient-encoded localized spectroscopy at this field strength.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times
0 Comments

More articles like this

WET, aT1- andB1-Insensitive Water-Suppression Meth...

on Journal of Magnetic Resonance... Jan 01, 1994

Assessment of metabolic fluxes in the mouse brain...

on Journal of cerebral blood flow... May 2015

Ab initio molecular orbital study of structure and...

on Chemical Physics Letters Jan 01, 1993

1H NMR of intact muscle at 11 T.

on FEBS Letters Jan 09, 1984
More articles like this..