Affordable Access

Publisher Website

Constancy check of beam quality in conventional diagnostic X-ray equipment

Authors
Journal
Applied Radiation and Isotopes
0969-8043
Publisher
Elsevier
Publication Date
Volume
66
Issue
10
Identifiers
DOI: 10.1016/j.apradiso.2008.03.007
Keywords
  • Ionization Chambers
  • X-Radiation
  • Quality Control
  • Diagnostic Radiology
  • Tandem System
Disciplines
  • Medicine

Abstract

Abstract A tandem ionization chamber was developed for quality control programs of X-ray equipment used in conventional radiography and mammography. A methodology for the use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities was established. The application at a medical X-ray imaging facility of this established methodology is presented. The use of the tandem chamber in the constancy check of diagnostic X-ray beam qualities is a useful method to control the performance of the X-ray equipment.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times
0 Comments

More articles like this

Constancy check of beam quality in conventional di...

on Applied Radiation and Isotopes Jan 01, 2008

Checking diagnostic x-ray beam quality.

on British Journal of Radiology March 1968

[Control of x-ray equipment for quality assurance...

on Röntgenpraxis; Zeitschrift fu... July 1988
More articles like this..