Abstract

The results indicate that the helical field deliveries in this study (including two clinical fields) do not introduce changes on the ion chamber correction factors for dosimetry. For those two specific clinical cases, ratios of chamber readings accurately represent field output factors. The values observed here for intermediate calibration fields are in agreement with previously published data based on alanine dosimetry but differ from values recently reported obtained via radiochromic dosimetry.