Comparison of four commercially available 81Rb-81mKr generators: quality testing using a breathing lung phantom.

Authors
  • Camps, J A
  • Blokland, J A
  • Blok, D
  • Pauwels, E K
Type
Published Article
Journal
International Journal of Radiation Applications and Instrumentation Part B Nuclear Medicine and Biology
Publisher
Elsevier
Publication Date
Jan 01, 1990
Volume
17
Issue
3
Pages
269–272
Identifiers
PMID: 2341281
Source
Medline
License
Unknown

Abstract

Four commercially available 81Rb-81mKr generators, including their elution systems, were compared. Considerable differences in useful output between the various generators were found. When a standardized elution system was used, the differences became much less. Differences in the tubing and inaccuracy of the flow meters were found to be the major causes of the original differences. The user should be aware of a loss of output that may occur when the generator is eluted with dry air.