Clinical field site evaluation of the FACSCount for absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell count determinations in Thailand.

Authors
  • N L Young
  • P Ponglertnapakorn
  • N Shaffer
  • K Srisak
  • T Chaowanachan
  • V On-Thern
  • C Kittinunvorakoon
  • A Bunwattanakul
  • S Suksaweang
  • V Pobkeeree
  • J Punnotok
  • T D Mastro
Publication Date
Nov 01, 1997
Source
PMC
Keywords
License
Unknown

Abstract

The FACSCount flow cytometer absolute CD3+, CD3+ CD4+, and CD3+ CD8+ cell counts measured at a field site hospital laboratory in Thailand were compared to FACScan absolute counts obtained at a nearby research laboratory. Correlation coefficients for 208 samples were > or = 0.95. The FACSCount was accurate, and it was easier and less expensive to operate than the FACScan. Additionally, the FACScan-generated lymphocyte percentage value was accurate for use with the FACScan SimulSET software.