Affordable Access

Publisher Website

Self-Retaining 27-Gauge Transconjunctival Chandelier Endoillumination for Panoramic Viewing During Vitreous Surgery

Authors
Journal
American Journal of Ophthalmology
0002-9394
Publisher
Elsevier
Publication Date
Volume
143
Issue
1
Identifiers
DOI: 10.1016/j.ajo.2006.07.051
Disciplines
  • Design
  • Medicine

Abstract

Purpose To report the development of a 27-gauge self-retaining transconjunctival chandelier endoilluminator for panoramic viewing during vitrectomy. Design New surgical instrument. Methods The tip of the illuminating fiber is 27-gauge (0.35 mm) and shaped like a cone to provide wide-angle illumination. The tip, inserted approximately 3 mm into the vitreous cavity transconjunctivally, provides diffuse illumination. The optical fiber covered by a malleable sleeve can be retained in the eyeball without a suture. Results The slim light fiber design (0.35 mm) stabilizes the 27-gauge tip in the eye and provides up to 25 lumens of measured illumination. The adequate lighting potential and self-retaining design allow wide-angle visualization and bimanual manipulation in challenging cases. Easy insertion and removal without suture placement make it convenient to change the position of the device during surgery and efficiently shorten surgical time. Conclusions This suture-free 27-gauge endoillumination chandelier improves the efficacy and efficiency of vitreous surgery.