VMAT testing for an Elekta accelerator

Authors
Publisher
Multimed Inc.
Publication Date
Keywords
  • Medical Physics
  • Vmat Commissioning
  • Imat Commissioning
  • Linac Quality Assurance
  • Dmlc
  • 87.55.De
  • 87.55.Qr
  • 87.56.Bd
Disciplines
  • Medicine

Abstract

Volumetric-modulated arc therapy (VMAT) has been shown to be able to deliver plans equivalent to intensity-modulated radiation therapy (IMRT) in a fraction of the treatment time. This improvement is important for patient immobilization/localization compliance due to comfort and treatment duration, as well as patient throughput. Previous authors have suggested commissioning methods for this modality. Here, we extend the methods reported for the Varian RapidArc system (which tested individual system components) to the Elekta linear accelerator, using custom files built using the Elekta iComCAT software. We also extend the method reported for VMAT commissioning of the Elekta accelerator by verifying maximum values of parameters (gantry speed, multileaf collimator (MLC) speed, and backup jaw speed), investigating: 1) beam profiles as a function of dose rate during an arc, 2) over/under dosing due to MLC reversals, and 3) over/under dosing at changing dose rate junctions. Equations for construction of the iComCAT files are given. Results indicate that the beam profile for lower dose rates varies less than 3% from that of the maximum dose rate, with no difference during an arc. The gantry, MLC, and backup jaw maximum speed are internally consistent. The monitor unit chamber is stable over the MUs and gantry movement conditions expected. MLC movement and position during VMAT delivery are within IMRT tolerances. Dose rate, gantry speed, and MLC speed are accurately controlled. Over/under dosing at junctions of MLC reversals or dose rate changes are within clinical acceptability.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times
0 Comments

More articles like this

VMAT testing for an Elekta accelerator.

on Journal of applied clinical me... 2012

MLC tracking for Elekta VMAT: a modelling study.

on Physics in Medicine and Biolog... Dec 07, 2011

[Investigation of Elekta linac characteristics for...

on Zhongguo yi liao qi xie za zhi... January 2012

Fixed bending current for Elekta SL25 linear accel...

on Journal of medical engineering... 2001
More articles like this..