Analysis of a discrepancy in electron–beam dose comparison between chemical dosimeters and a calorimeter

Authors
Journal
Applied Radiation and Isotopes
0969-8043
Publisher
Elsevier
Publication Date
Volume
55
Issue
1
Identifiers
DOI: 10.1016/s0969-8043(00)00373-0
Keywords
  • Electron Beam
  • Absorbed Dose
  • Depth-Dose Distribution
  • Sandwich Phantom
  • Semiempirical Code
Disciplines
  • Chemistry

Abstract

Abstract A 2.1% discrepancy was reported from an electron-beam dose comparison between liquid chemical dosimeters and a water calorimeter. Two methods, depth–dose measurement in a sandwich phantom and the use of a semiempirical depth–dose code EDMULT, were used to analyze the depth-dose distribution in the dosimeters and the calorimeter and to calculate their dose ratio. Results have shown that the discrepancy can be explained by the difference in the effective detection volume.