The automation of sediment urinalysis using a new urine flow cytometer (UF-100)

Authors
  • Fenili, D
  • Pirovano, B
Type
Published Article
Journal
Clinical chemistry and laboratory medicine : CCLM / FESCC
Publication Date
Dec 01, 1998
Volume
36
Issue
12
Pages
909–917
Identifiers
PMID: 9915222
Source
Medline
License
Unknown

Abstract

The UF-100 analyser is a fully automated instrument that stains the DNA and the membranes of the formed elements in native urine. The sample then passes as a laminar flow through a laser beam and light scattering, fluorescence and impedance are measured. The main purpose of the present work was to assess the analytical performance and the accuracy of the measurements of the UF-100 analyser. No carryover was observed, while the linearity was higher then the upper limit (40000 total particles microl(-1)) suggested by the manufacturer. The within-run imprecision was low, ranging from 17.7 % to 2.4% and was up to threefold better than manual microscopy. Comparison of results obtained by sediment microscopy (performed according to National Committee for Clinical Laboratory Standards (NCCLS) recommendations) and by the UF-100 analyser showed a linear correlation with r = 0.833 for erythrocytes, r = 0.934 for leukocytes, r = 0.880 for epithelial cells and r = 0.40 for casts. To evaluate the reliability of the UF-100 analyser in detecting bacteria we compared the results with the microbial culture (n = 608). Using a cut-off value of bacterial count above 1800 degrees l(-1) and at leukocyte count above 45 microl(-1), the analyser detected positive cultures with a sensitivity of 87 % and a specificity of 80 %. In conclusion, the UF-100 analyser can improve the work flow, increasing the output of urinalysis by reducing the number of specimens submitted for microscopy. Also the method provides reliable information for the identification of urinary tract inflammation and bacterial infection.

Report this publication

Statistics

Seen <100 times