Analysis of the mechanical impedance of bone-anchored hearing aids.

Authors
Type
Published Article
Journal
Acta oto-laryngologica
Publication Date
Volume
89
Issue
1-2
Pages
85–92
Identifiers
PMID: 7405582
Source
Medline

Abstract

Some patients who need hearing aids are unable to use an apparatus which transmits the sound via the external ear canal and have to use a bone conduction hearing aid. The bone vibration transducer of this aid is applied to the skin over the mastoid process and the sound is transmitted via the soft tissue and bone to the cochlea. The pressure needed to apply the transducer often gives the patient discomfort and the damping effect of the soft tissue gives poor quality of the sound transmitted. Advances in the ability to permanently implant foreign material in the body and perform permanent skin penetration has made it possible to develop a bone-anchored hearing aid. Fourteen patients have been equipped with such hearing aids. To be able to give these patients the best hearing aid, a new transducer has to be constructed to match the new situation. The impedance of the bone-anchored titaniumscrew/skull has been studied and the resistance and reactance of the mechanical impedance have been measured. The influence of a damping soft tissue layer over the bone has been analyzed. The difference between the impedance of the skull and the impedance of the soft tissue + skull was in the order of 10 to 25 dB depending on the frequency.