A low-cost, MR-compatible olfactometer

Authors
Publication Date
Source
PMC
Keywords
  • Article
Disciplines
  • Design

Abstract

We present a design for an olfactometer, suitable for fMRI experiments, that can be constructed at extremely low cost. The olfactometer presents odors directly to the nose via a nasal cannula at unobtrusively low flow velocities, with no large assemblies required on or near the subject's face. The olfactometer can be controlled manually, or by computer via a serial interface. A validation study verified that the olfactometer reliably presents odors to test subjects. Errors and response latency times decreased with increased flow rate in an orderly manner, as expected.

There are no comments yet on this publication. Be the first to share your thoughts.