Acoustic method for calibration of audiometric bone vibrators. II. Harmonic distortion.

Authors
  • Ginter, Samantha M
  • Margolis, Robert H
Type
Published Article
Journal
The Journal of the Acoustical Society of America
Publication Date
Jul 01, 2013
Volume
134
Issue
1
Identifiers
DOI: 10.1121/1.4804944
PMID: 23862903
Source
Medline
License
Unknown

Abstract

A previous report [Margolis and Stiepan (2012). "Acoustic method for calibration of audiometric bone vibrators," J. Acoust. Soc. Am. 131, 1221-1225] described a reliable, inexpensive, acoustic method for calibration of audiometric bone vibrators. As a follow up to that report harmonic distortion measurements were made with the standard electromechanical method and the acoustic method using five Radioear B71 vibrators and one Radioear B81 prototype vibrator. Lower distortion was seen for measurements made with the acoustic method compared to the electromechanical method and for the Radioear B81 vibrator compared to the Radioear B71 vibrator.