Affordable Access

Publisher Website

Analysis of a bomb calorimeter failure

Authors
Journal
Journal of Chemical Health and Safety
1871-5532
Publisher
Elsevier
Publication Date
Volume
17
Issue
2
Identifiers
DOI: 10.1016/j.jchas.2009.08.002

Abstract

Shrapnel and the release of hot, pressurized liquid occurred during a routine bomb calorimetry experiment. Corrosion was found to have weakened the calorimeter bomb's valve assembly, resulting in shearing of the assembly during the experiment.