Affordable Access

Publisher Website

Modification of commerical and custom-built slab gel electrophoresis units for two-dimensional polyacrylamide gel electrophoresis

Authors
Journal
Analytical Biochemistry
0003-2697
Publisher
Elsevier
Publication Date
Volume
89
Issue
2
Identifiers
DOI: 10.1016/0003-2697(78)90370-6
Disciplines
  • Biology

Abstract

Abstract Many commercial and custom-built slab gel electrophoresis units can be modified to function as two-dimensional polyacrylamide gel electrophoresis units with the insertion of Plexiglas adapters. These adapters can be made for about $50 a pair and can be used for either temporary or permanent modification of the slab gel units. The physical dimensions of the adapters can be varied to permit great flexibility in the diameter of cylinder gels and the thickness of slab gels that can be run together. For example, proteins from 6-mm cylinder gels can be easily separated on 1-mm slab gels, which can then be dried for autoradiography.