Affordable Access

Microbial biosensor for the analysis of 2,4-dichlorophenol

Authors
Publisher
Taylor & Francis
Publication Date
Keywords
  • Chemistry

Abstract

A flow injection cell-based biosensor was constructed for 2,4-dichlorophenol (DCP) analysis by using a Clark-type oxygen electrode as a transducer. A mixed bacterial culture capable to biodegrade DCP was immobilized between a Teflon membrane and a dialysis membrane and attached to the oxygen electrode. Optimization of the flow rate, the injection volume, the carrier buffer concentration, and pH was carried out. Under optimum conditions (100 mM phosphate buffer, pH 7.50; flow rate 0.10 mL min(-1); sample volume 100 mL), the sensor response was linear between 0.01 and 0.30 mM DCP. The detection limit was 0.02 mM DCP, and the sensor was quite stable during 5 days of operation.

There are no comments yet on this publication. Be the first to share your thoughts.

Statistics

Seen <100 times
0 Comments

More articles like this

Amperometric glucose biosensor based on NiFe2O4nan...

on Sensors and Actuators B Chemic... Jan 01, 2010

Raman analysis of C2H4molecular jets excited by a...

on Chemical Physics Letters Jan 01, 1984

Extractability of microbial14C and15N following ad...

on Soil Biology and Biochemistry Jan 01, 1990
More articles like this..