Monte Carlo simulation of x-ray fluorescence analysis of gold in kidney using99mTc radiopharmaceutical

Authors
Journal
Radiation Physics and Chemistry
0969-806X
Publisher
Elsevier
Volume
91
Identifiers
DOI: 10.1016/j.radphyschem.2013.05.027
Keywords
  • Xrf Analysis
  • Kidney
  • Gold
  • Skin Dose
Disciplines
  • Design

Abstract

Abstract The objective of this study was to design a system in order to analyze gold and other heavy elements in internal organs using in vivo x-ray fluorescence (XRF) analysis. Monte Carlo N Particle code MCNP was used to simulate phantoms and sources. A source of 99mTc was simulated in kidney to excite the gold x-rays. Changes in K XRF response due to variations in tissue thickness overlying the kidney at the measurement site were investigated. Different simulations having tissue thicknesses of 20, 30, 40, 50 and 60mm were performed. Kα1 and Kα2 for all depths were measured. The linearity of the XRF system was also studied by increasing the gold concentration in the kidney phantom from 0 to 500µgg−1 kidney tissue. The results show that gold concentration between 3 and 10µgg−1 kidney tissue can be detected for distance between the skin and the kidney surface of 20–60mm. The study also made a comparison between the skin doses for the source outside and inside the phantom.