Abstract

We present methods to improve the energy resolution of single channel, room-temperature Cadmium-Zinc-Telluride (CZT) detectors. A new preamplifier design enables the acquisition of the actual transient current from the crystals and straightforward data analysis methods yield unprecedented energy resolution for our test-detectors. These consist of an eV-CAPture Plus crystal as standard and 1 cm cube Frisch collar crystals created in-house from low-grade coplanar grid detectors. Energy resolutions of 1.9% for our collar detectors and 0.8% for the eV crystal at 662 keV were obtained. The latter compares favourably to the best existing energy resolution results from pixel detectors.

There are no comments yet on this publication. Be the first to share your thoughts.