Affordable Access

Publisher Website

Performance of the new high precision luminosity monitor of DELPHI

Authors
Journal
Nuclear Physics B - Proceedings Supplements
0920-5632
Publisher
Elsevier
Publication Date
Volume
54
Issue
3
Identifiers
DOI: 10.1016/s0920-5632(97)00122-9

Abstract

The STIC calorimeter was installed in the DELPHI detector in 1994. The main goal is to measure the luminosity with an accuracy better than 0.1%. The calorimeter was built using the “Shashlik” technique. The light is collected by wavelength shifting fibers and readout by phototetrodes that can operate inside the magnetic field. The detector performance during the 1994–1995 data taking is presented. The different contributions to the systematic error on the luminosity measurement are discussed.