An evaluation of the Auto Analyzer SMA-4

Authors
Publication Date
Source
PMC
Keywords
  • Articles

Abstract

The Technicon AutoAnalyzer SMA-4 has been systematically evaluated. Carry-over from the sample cups was found to be within an acceptable range. The precision and accuracy of the four parameters determined by the instrument have been investigated. The haemoglobin results were found to be accurate. There was good agreement with leucocyte counts performed on the model A Coulter electronic cell counter. After certain modifications had been made in the manifold, satisfactory degrees of accuracy were also obtained for the erythrocyte counts. Although the determination of the haematocrit by conductance is influenced by a variety of factors, the mean coefficient of variation was found to be 0·83% and 95% of the results agreed within ± 2% of those obtained by the microhaematocrit centrifugation method.